Rapid, Simple, Accurate Mycoplasma Detection to Anyone & Anywhere
The AdvSHENTEK Mycoplasma DetectInnova System transformed RMM (rapid microbiological method) into a streamlined, fast, fully automated, and easily portable laboratory system. It integrates nucleic acid extraction and real-time PCR assay, allowing a turn-around time of 2.5 hours from sample to result. Importantly, it only requires less than 5 minutes of hands-on time without specific DNA extraction and PCR expertise to operate.
Easy on-time testing:
Complete automation from sample to result;
Hands-on time < 5 mins;
High Sensitivity & Efficiency:
LOD < 10 CFU/mL;
4 standardized assays in a single run;
Accessible Anywhere &Anyone:
No PCR lab or skills required;
Seamless integration from extraction to amplification without manual intervention;
Data Reliability:
Fully validated workflow;
Effective internal control across every stage;
21 CFR Part 11 compliance;
High Specificity:
No cross-reactivity with different bacteria or cell lines;
No sample matrix interference;
Comprehensive coverage:
Able to detect over 180 species of Mycoplasma, Spiroplasma and Acholeplasma;
Through intensive bioinformatics analysis of the amplification targets, the detection system allows genomic DNA of nearly 200 species, including 135 species of Mycoplasma, 38 species of Spiroplasma, and 8 species of Acholeplasma.
No cross-reactivity with Clostridium, Lactobacillus, Streptococcus, and different cell lines.
Table 3 Results of Cross-Reactivity
Cross reactant NO. | Resource | Cross Reactivity | Cross reactant NO. | Resource | Cross Reactivity | ||
1 | CHO | N/A | 10 | Pichia pastoris | N/A | ||
2 | Vero | N/A | 11 | Saccharomyces cerevisiae | N/A | ||
3 | 293T | N/A | 12 | Hans enula polymorpha | N/A | ||
4 | HEK293 | N/A | 13 | Clostridium acetobutylicum | N/A | ||
5 | MDCK | N/A | 14 | Clostridium perfringens | N/A | ||
6 | NS0 | N/A | 15 | Clostridium sporogenes | N/A | ||
7 | Sf9 | N/A | 16 | Streptococcus pneumoniae | N/A | ||
8 | Hi5 | N/A | 17 | Streptococcus agalactiae | N/A | ||
9 | E.coli | N/A | 18 | Streptococcus mutans | N/A | ||
Table 4 Results of Sample Matrix Interference
List of culture media, serum, and cryoprotectants tested do not interfere with mycoplasma detection using the SHENTEK DetectInnova system.
NO. | Sample Matrix | Signal | |||||
1 | ExpiCHOTM | N/A | |||||
2 | ExpiCHOTM(10%DMSO) | N/A | |||||
3 | DMEM(10%FBS) | N/A | |||||
4 | DMEM(20%FBS+10%DMSO) | N/A |
Table 5 Matrix robustness results
Strain | Matrix | Tset Conc. (CFU/mL) | Detection rate(%) | Acceptance Criteria | Conclusion | ||
Mycoplasma orale | CHO-S (10⁷cells/mL) | 10 | 95.8 | Detection rate at least over 95%. | Pass | ||
Mycoplasma hyorhinis | 10 | 100 | Pass | ||||
Mycoplasma pneumoniae | 10 | 100 | Pass | ||||
Mycoplasma fermentans | 10 | 100 | Pass | ||||